Measured value characteristics
Sensitivity drift
< 2% of the respective full scale value per week
Zero point drift
< 2% of the respective full scale value per week
Span drift
< 2% of the respective full scale value per week
Setting time t
90
< 200 s, total measuring path as from probe extraction
13.2.2
Ambient conditions
Table 13: Operation
Ambient conditions in operation
Installation location
Indoor installation
Ambient temperature
+10 ... +40 °C
Relative humidity
< 90% (without condensate)
Air pressure
850 ... 1100 hPa
Table 14: Storage
Ambient conditions in storage
Ambient temperature
-20 ... +70 °C
Relative humidity
< 90% (without condensate)
13.2.3
Housing
Table 15: Design
Design
Design
Installation plate
Material, general
Steel plate, aluminium cast
Dimensions
Installation
Installation in analyzer enclosure or system container
Weight
Approx. 200 kg
Materials with media contact
•
Nickel-plated aluminium
•
PTFE
•
Stainless steel 1.4401
Degree of protection
Depending on analyzer enclosure / system container
used
13.2.4
Interfaces and protocols
Table 16: Interfaces and protocols
Operation and interfaces
Operation
Via display or Google Chrome browser with SOPASair
software, several operating levels, password-protected
Display and input
Foiled color display with touchscreen
Analog inputs/outputs
Optional
Digital inputs/outputs
Optional
Data interface
1 x Modbus TCP/IP
Profibus
Configurable
Profinet
Configurable
Remote support
SICK MPR (optional)
PC operation
Browser Google Chrome with SOPASair via Ethernet
13
TECHNICAL DATA
56
O P E R A T I N G I N S T R U C T I O N S | MCS200HW-MP
8026608/AE00/V1-0/2021-10 | SICK
Subject to change without notice