Shining 3D Transcan-C_User Manual_V1.4.2
19
Functional description.
1: Left and right camera windows.
2: Location and placement of calibration board.
3: Operation prompts.
4: Scanning range. Select 150mm or 300mm on the device adjustment interface.
4.2.1.
Camera Calibration
The calibration board needs to be placed in ten positions during camera calibration, and the
positions are operated according to the software wizard.
Place the calibration board according to the
software wizard prompts.
Click on the camera brightness scale and adjust until the camera windows turn slightly red.
Place the calibration board according to the following figure. Ensure that the placed position is
consistent with that shown in the figure.
The calibration board can be divided into upper, lower, left and right. In the first step, make sure
that the five big circles in the red box are placed near the bottom of the calibration pad, as shown in
the figure below.
Ensure that the calibration board is placed smoothly and facing the probe. Click
Capture
. Do
not move the calibration board during the collection process.
After the collection is completed, the software interface displays for the second step calibration.
The calibration board is divided into upper, lower, left and right. Please make sure that the first
step is as follows: