Shining 3D Transcan-C_User Manual_V1.4.2
16
8)
Loosen the screws on the left and right sides of the camera, move the camera to the proper
position. After the adjustment is completed, tighten the screws.
9)
Face the scanner with white side of the calibration board. Use the
Allen key
placed under the
camera to loosen the screws at the rear of the camera, adjust the left and right angles of the camera
so that the intersection of the black cross is basically on the red vertical line of the camera window,
and then tighten the screws.
150mm
300mm
Allen key specific location
Two sizes of Allen keys
10)
Unscrew the screw on the aperture and turn the aperture to adjust the brightness of the left and
right camera windows. The recommended value is 5.6. When the adjustment is complete, tighten the
screw (or align white side of the calibration plate with the scanner and adjust the camera brightness
to the highest brightness that does not show large areas of red).