Shining 3D Transcan-C_User Manual_V1.4.2
15
5)
Roughly adjust the height of the equipment, make the light projected on the center of the
calibration board.
6)
150mm: Ensure the projected area aligns with the left and right inner edge of the rubber cover
and the projected cross should aim at the center point. The scanner aiming direction should be
perpendicular to the calibration board.
300mm: Ensure the top and bottom edges of projected area aligns with two rows of points on
the calibration board and the projected cross should aim at the center point. The scanner aiming
direction should be perpendicular to the calibration board.
150mm
300mm
7)
Face the scanner with white side of the calibration board and turn the outer ring of the optical
machine to adjust the black cross to a clear and optimal state.
150mm
300m