Page 44 of 60
SGM-101F - programming menu
10.64 M64 - AL3 analog input programming
NOT AVAILABLE.
AI3 Value Range
20 ~ 100
10.65 M65 - AL4 analog input programming
NOT AVAILABLE.
AI4 Value Range
20 ~ 100
10.66 M66 - AL5 analog input programming
NOT AVAILABLE.
AI5 Value Range
0 ~ 6
10.67 M67 - Frequency output range programming
In this menu it’s possible to set the OCT output range when it set as a
frequency output proportional to the measured instantaneous
fl
ow rate.
FO Frequency Range
0 ~ 1000 Hz
10.68 M68 - Frequency output low fl ow rate programming
In this menu it’s possible to set the frequency output low
fl
ow rate.
Low FO Flow Rate
0
m3/h
10.69 M69 - Frequency output high fl ow rate programming
In this menu it’s possible to set the frequency output high
fl
ow rate.
High FO Flow Rate
10800 m3/h
10.70 M70 - Backlight interval programming
In this menu it’s possible to set the display backlight time.
Range: 0÷60000 seconds.
LCD Backlight Option
10
Sec
10.71 M71 - LCD contrast programming
In this menu it’s possible to set the LCD contrast.
Range: 00÷31
Press “ENTER” to enter, then increase the contrast with the
“UP ARROW” or decrease the contrast with “SCROLL”.
Press “ENTER” to store.
LCD Contrast
18
Working Timer
00000175:42:15
10.72 M72 - Operation time displaying
In this menu it’s possible to display the SGM-101F operation total time,
since the last timer reset.
To reset the timer, proceed as follows:
Press “ENTER 2 times to enter, then select “YES” with
“UP ARROW” and press “ENTER” to con
fi
rm the reset.
Reset Working Timer
>
YES
10.73 M73 - #1 Q min. alarm programming
In this menu it’s possible to set the minimum
fl
ow threshold for the #1
alarm.
1# Alarm LOW Value
0
m3/h