SGM-101F - programming menu
10.23 M23 - Transducers type programming
The display shows the previous setting.
The available models are (catalog models highlighted in bold):
0. Standard-M
1. Insertion Type C
2. Standard-S
3. User Type (some additional menus are activated for the
non-standard transducers characteristics when 3 option is selected)
4. Standard-B
5. Insertion B(45)
6. Standard-L
7. JH-Polysonics
8. Standard-HS
9. Standard-HM
10. Standard-M1
11. Standard-S1
12. Standard-L1
13. PI-Type
14. FS410 (FUJI)
15. FS510 (FUJI)
16. Clamp-on TM-1 (see features on page 8)
17. Insertion TC-1 (see features on page 9)
18. Clamp-on TS-1
19. Clamp-on TS-2 (see features on page 7)
20. Clamp-on TL-1 (see features on page 8)
21. Insertion TLC2 (see features on page 9)
22. Clamp-on M2
23. Clamp-on L2
Transducer Type
16. Clamp-on TM-1
10.24 M24 - Transducers mounting method programming
The display shows the previous setting.
The available mounting methods are:
0. V
1. Z
Transducer Mounting
0.
V
10.25 M25 - Transducers mounting distance displaying
The display shows the automatically calculated transducers mounting
distance.
Transducer Spacing
34.334mm
10.26 M26 - Data storage mode programming
The display shows the previous setting.
Available settings:
0. Use RAM Settings (recommended); all the latest programming
changes are automatically stored in the EEPROM
1.Solidify Setting; to store in the EEPROM the latest changes to
any menu, must go back to M26 and press “ENTER”.
Default Setting
1. Solidify Setting