Page 48 of 60
SGM-101F - programming menu
10.88 M88 - Heat meter totalizer multiplier programming
The display shows the previous setting.
Default value: x1
Available settings:
0.
x0.0001 (E-4)
1.
x0.001 (1E-3)
2.
x0.01
3.
x0.1
4.
x1
5.
x10
6.
x100
7.
x1000
8.
x10000 (E4)
9.
x100000 (E5)
10.
x1000000
(E6)
Energy Multiplier
4. X1
(E0)
10.89 M89 - Temperature differential displaying
In this menu it’s possible to display the temperature difference between
supply and return.
Temperature Diff.
0.0039
C
10.90 M90 - Transducers signal power and quality displaying
In this menu it’s possible to display the ultrasonic transducers ef
fi
ciency
(UP and DN) and the ultrasonic signals quality (Q) processed by
SGM-101F. For the “UP” (upstream transducer) and “DN” (downstream
transducer) the 00.0 value indicates the ultrasonic signal non-reception,
while the 99.9 value indicates the ultrasonic signal excellent reception;
Normally the value is greater than 60.0. The processed ultrasonic signals
quality (Q), has a range from 00.0 to 99.9.
Normally the “Q” value is greater than 60.0.
St Quality
UP:78.5 DN:78.8 Q=92
10.91 M91 - TOM/TOS % displaying
In this menu it’s possible to display the ratio between the calculated and
the measured transit time. Normally the value should be 100 ±3%.
Differences in excess of the above, could mean improper transducers
mounting, or incorrect programming values.
T O M / T O S
3.9478 %
10.92 M92 - Sound velocity in the fl uid displaying
In this menu it’s possible to display the sound speed in the
fl
uid, measured
by the SGM-101F. Normally the value should be similar to what is set in
M21, accessible when M20 is set to “Other Liquid”.
A signi
fi
cant values difference, could mean improper transducers mounting,
or incorrect M21 programming values.
Fluid Sound Velocity
1486.35 m/s
10.93 M93 - Flight time and delta T.
In this menu it’s possible to display the
fl
ight time measured by the
SGM-101F and the difference in
fl
ight times, UP - DN.
TotalTime, DeltaTime
624.72uS 251.67nS
10.94 M94 - Reynolds number displaying
In this menu it’s possible to display the calculated Reynolds number value.
Reynolds No, Profile
12354.8
0.97563