Page 16 of 60
SGM-101F -
installation
7-INSTALLATION
7.1 MEASURING POINT SELECTION
The transducers must be mounted on a pipe section which allows to respect the minimum distance between
the element of resistance to fl ow, such as curves or derivations, and the measuring point.
See the following table.
Flow resistance element
Upstream side
Downstream side
90° curves
15 x DN
5 x DN
T junction
40 x DN
5 x DN
Adaptors
18 x DN
5 x DN
Valves
40 x DN
5 x DN
Pumps
20 x DN
P