7.6 INSERTION TRANSDUCER INSTALLATION
Steps required for proper installation:
1 - with encased pipe, check that there is the minimum space required for the transducers installation.
2 - with encased pipe, check that the free section length of the pipe is the minimum required.
540
Min. distance
from wall
540
Min. distance
from wall
ø Pipe
Distance
transducer
Distance between the tran 200mm.
Transducer
Pipe
Transducer
3 - Procuring the necessary equipment for drilling the pressure pipes .
4 - Set the pipe parameters: in the menu M23 choose the option 17 or 21 (TC-1 or TLC-2 insertion transducer);
in the menu M24 choose 1. (Z installation) and in the menu M25 check the positioning distance.
5 - Determine the best location for installation on pipe .
6 - Installing the ball valve base.
Metallic pipe
Non-metallic pipe
Valve base
welded to the pipe
Metal band