The transducers positioning points to be avoided are:
- vertical pipes with the downward
fl
ow, because they may not be completely
fi
lled with
fl
uid.
- inclined pipes with the downward
fl
ow, because they may not be completely
fi
lled with
fl
uid.
- the transducers must never be placed in the highest point of the concerned hydraulic circuit, because there
is greater chance that in that pipeline section will create air pockets.
- vertical open drain pipes without a section restriction to avoid sudden pipe emptying during
fl
ow measurement.
7.2 POSITIONING DISTANCE
The value (calculated automatically by the system) shown in menu M25 refers to the “Lout” mounting distance
between the two transducers, as shown in the following
fi
gures.
Lout
Lout