9 – Fault finding
MultiExact Gas Analyser
Message
Measurement
screen icon
Recommended actions
Pressure calibrate
Raised during calibration of the pressure
transducer (see Section 6.3).
Relay not detected
Contact Servomex or your local Servomex agent
for assistance.
Relays overridden
Raised when the alarm and status relays have
been overridden (see Section 7.1).
Remote activation
Raised when the Service in Progress status is
remotely activated via the SIP relays on the
Calibration Card (see Section 5.2.4).
Rem lo cal denied
Raised when a remote low calibration cannot
execute because the current reading is outside
tolerances (see Section 5.5.3).
Rem hi cal denied
Raised when a remote high calibration cannot
execute because the current reading is outside
tolerances (see Section 5.5.3).
Static RAM fault
Turn the analyser off, and then turn it on again. If
the fault message is then displayed again, contact
Servomex or your local Servomex agent for
assistance.
Sw IP not detected
Contact Servomex or your local Servomex agent
for assistance.
Temperature fault
Reduce temperature to within environmental limits
or contact Servomex or your local Servomex agent
for assistance.
Transducer error
Ensure that you are using the analyser in the
specified operating conditions (refer to Section 2).
If the fault persists, contact Servomex or your local
Servomex agent for assistance.
Transducer warming
Infra-red only, transducer is warming after power
on, measurement should be treated with caution
while status is active.
Tx calibrate
Raised during manual, remote or auto calibration of
the transducer.
88
05410001A / Revision 6