Appendix
MultiExact Gas Analyser
A9.9 Performance data: TCD
☞
The display indication given below is the default indication. You can configure
the analyser to provide other display indications (see Section 5.10).
☞
Performance data has been determined in accordance with EN61207.
Display indication
%
Analogue output range
Resolution
0.01%
Repeatability
0.5% of range
Linearity
± 1% of range
Intrinsic error (accuracy)
± 1% of range
Drift per month
± 1% of range
Output fluctuation
± 0.5% of range
Response time *
T90 15 seconds *
Flow effect
< 0.1%
#
Temperature coefficient
± 1% of range per 10 °C
* for 150 ml min
-1
.
Flow driven.100 – 200 ml min
-1
or
pressure driven 2 – 8 psig.
#
Over defined flow range
Gas and Ranges available
0
-1%
0
-2%
0
-5%
0
-10%
0
-20%
0
-30%
0
-50%
0
-100%
80
-100%
90
-100%
Ar in N
2
Ar in O
2
N
2
in Ar
He in N
2
H
e
in
O
2
134
05410001A / Revision 6