5 – Installation and set-up
MultiExact Gas Analyser
5.9.2 Entering a cross-interference (X-Interference) compensation
☞
You can only apply cross-interference compensation to O
2
purity or control
measurements.
☞
X-Interference compensation is disabled during calibration, and is not applied
to the values shown in Figure 74. All other outputs (that is, serial or mA
outputs) remain compensated.
Use the following procedure to enter a compensation to correct for an O
2
(oxygen)
measurement error:
1.
Select
➞
Set up
➞
X-Interference.
☞
If your analyser is configured for a single sample measurement, the
X-Interference offset screen is then displayed. Continue at step 3.
2.
With the X-Interference select screen
displayed select the required
measurement, then press the
soft key.
3.
The offset value shown on the
X-Interference offset screen is the
correction which will be applied to O
2
(oxygen) sample measurements
before they are displayed (or output).
4.
Edit the displayed offset as described
in Section 4.10.
Figure 59 – The X-Interference
offset screen
5.10
Selecting display units
You can change the measurement units shown on the display (and output). The
following display units are supported:
Units
Meaning
%
volume %
ppm
parts per million
vpm
volume parts per million
mg/m
3
mg m
-3
(milligrams per normal cubic metre)
mol/mol
mols per mol (or moles per mole)
% LEL
volume % of the Lower Explosive Limit