6 – Calibration
MultiExact Gas Analyser
Figure 81 – The auto val/cal
failed screen
6.2.3 Stopping an automatic validation/calibration
☞
If you stop an automatic validation/calibration after the pre-warning phase, the
inerting phase (if active) and the final flushing phase will always be completed.
To stop an automatic validation/calibration which is in progress, close the stop
auto val switch input on the appropriate interface connection (see Section 5.2.5) or
use modbus command see Appendix A3. Alternatively, whenever an automatic
validation/calibration is in progress and the
(stop) icon is shown on the
auto val/cal status screen (see Figure 78 and Figure 79), you can press the
soft button to manually stop the automatic validation/calibration.
When you press the
soft button, the
auto val/cal stop confirmation screen is
shown (see Figure 82).
Press the
soft key to stop the
automatic validation/calibration. The
measurement screen (Figure 3) will then
be displayed again.
Figure 82 – The auto val/cal
stop confirmation screen
☞
Press the
soft key to continue with the automatic validation/calibration.