MultiExact Gas Analyser
5 – Installation and set-up
1.
Select
➞
Set up.
2.
Select the required transducer and
alarm.
Figure 63 – The alarm set up screen
3.
Select the required mode (none, low
or high), then press the
soft
key.
Figure 64 – The alarm mode screen
4.
Scroll up or down to edit the appropriate settings (using the method described
in Section 4.10):
•
Latching (Section 5.11.2)
•
Level (sets the gas concentration level at which the alarm is to trigger)
•
Hysteresis (Section 5.11.3).
Each measurement has a 'Follow' option:
•
If the 'Follow' option is set to 'No', the alarm will be inhibited during calibration.
•
If the 'Follow' option is set to 'Yes', the alarm will not be inhibited during calibration.
1.
Select
➞
Follow.
2.
Select the required alarm.
Figure 65 – The alarm follow screen
3.
Select the correct option: "Yes" or
"No".
Figure 66 – The alarm follow
options screen
05410001A / Revision 6
61