9 – Fault finding
MultiExact Gas Analyser
Message
Measurement
screen icon
Recommended actions
Communication fail
Turn the analyser off, and then turn it on again. If
the fault message is then displayed again, contact
Servomex or your local Servomex agent for
assistance.
Database fault
Turn the analyser off, and then turn it on again. If
the fault message is then displayed again, contact
Servomex or your local Servomex agent for
assistance.
Date/Time invalid
This usually occurs because the electrical supply to
the analyser has been switched off for more than a
week.
Switch on the electrical supply, then set the date/
time as described in Section 5.4.6. If the fault
persists, contact Servomex or your local Servomex
agent for assistance.
Fatal fault
Contact Servomex or your local Servomex agent
for assistance.
Flow calibrate
Raised during calibration of the flow alarm
(see Section 5.12).
Flow H/W fault
Contact Servomex or your local Servomex agent
for assistance.
Flow cal high diff
The zero and normal calibration points are too far
apart, check normal flow is within limits
(see Section 2.4) and re-calibrate the flow alarm.
Flow cal low diff
The zero and normal calibration points are too
close, check normal flow is within limits
(see Section 2.4) and there is no flow during zero
calibration. Recalibrate the flow alarm.
Flow temp fault
Contact Servomex or your local Servomex agent
for assistance.
Heater Fault
Contact Servomex or your local Servomex agent
for assistance.