Appendix
MultiExact Gas Analyser
Register
Name
Comments
1112
Communication fail
Transducer not responding.
1113
Incorrect transducer type
Incorrect type of transducer fitted.
1114
Transducer not detected
Transducer is unplugged or broken.
1115
Low calibration fail
Auto cal low failed.
1116
High calibration fail
Auto cal high failed.
1117
Low validation fail
Auto val low failed.
1118
High validation fail
Auto val high failed.
1119
Remote low cal denied
Remote low calibration denied.
1120
Remote high cal denied
Remote high calibration denied.
1121
mA fault
Hardware fault detected.
1122
mA not detected
Card is missing or the incorrect type
has been fitted.
1123
mA reset
A time out occurred on the milliamp
card.
1124
Volt fault
Hardware fault detected.
1125
Volt not detected
Card missing or the incorrect type has
been fitted.
1126
Volt reset
A time out occurred on the volt card.
1127
Pressure calibration
Pressure needs calibrating.
1128
Pressure fail
Hardware fault detected.
1129
Pressure low
Pressure low fault.
1130
Pressure high
Pressure high fault.
1131
Relay not detected
Card missing or incorrect type has
been fitted.
1132
Switch input not detected
Card missing or incorrect type has
been fitted.
1133
Flow fail
Hardware fault detected.
1134
Low flow alarm 1
Flow is below is below the low flow 1
alarm level
1135
Low flow alarm 2
Flow is below is below the low flow 2
alarm level
1136
Remote service in progress
Switch input has activated service in
progress.
110
05410001A / Revision 6