5 – Installation and set-up
MultiExact Gas Analyser
5.6
Calibration linking
☞
Calibration linking only applies to analysers configured for 2 measurements.
☞
If calibration is not linked, the alarms for each transducer will operate in
accordance with their corresponding 'During Calibration' options during
calibration.
Calibration linking is useful when you route a single gas stream into both inlets of the
dual measurement analyser.
If you link calibration, then:
•
When you calibrate transducer 1, the alarms (Section 5.11.1) and analogue output
(Section 5.7.2) for transducer 2 will operate in accordance with the transducer 2
'During Calibration' option during calibration.
•
When you calibrate transducer 2, the alarms and analogue output for transducer 1
will operate in accordance with the transducer 1 'During Calibration' option during
calibration.
1.
Select
➞
Set up
➞
Calibration.
2.
Select 'Yes' or 'No' for calibration
linking.
Figure 47 – The Calibration linking
screen
50
05410001A / Revision 6