5 – Installation and set-up
MultiExact Gas Analyser
3.
If you changed the pre-warning
setting from 'Off' to 'On', the
Auto val/cal
pre-warning
timer
screen will then be displayed
(see Figure 38). Edit the displayed
time as described in Section 4.10.
Figure 38 – The Auto val/cal
pre-warning timer screen
4.
Select 'On' or 'Off' for inerting.
Figure 39 – The Auto val/cal
inerting screen
5.
If you have changed the Inerting
setting from 'On' to 'Off', continue at
Step 9.
6.
Select the correct inert gas: 'Low' or
'High'.
Figure 40 – The Auto val/cal
inert gas screen
7.
Scroll to the auto val/cal inerting
timer screen, showing the currently
selected inerting time setting.
8.
Edit the time as described in
Section 4.10.
Figure 41 – The Auto val/cal
inerting timer screen
9.
Scroll to the auto val/cal flushing
screen, showing the currently
selected flushing time setting.
10.
Edit the time as described in
Section 4.10.
Figure 42 – The auto val/cal
flushing screen
48
05410001A / Revision 6