Tools & Setup
40
4.2.6 Odometer Calibration
This test ensures that the odometer is measuring distance accurately. Even though the
odometer is calibrated in the factory, you must calibrate it periodically to ensure accuracy. This
is particularly important as the calibration can change with different surface materials.
When selecting this option, you will see the screen in Figure 4-16. The following options exist:
Figure 4-16: Screen displaying current odometer calibration
•
Odometer Test
– This test verifies the odometer is calibrated properly. It involves
moving the system a set distance (5m or 1
0’, depending on units used), and comparing
the value obtained to a known value. The screen in Figure 4-17 will be displayed once
Odometer Test is pressed. Press
Start
to begin and then press
Finish
when you’ve
moved that distance. It will display a pass or fail, along with a percentage difference
from the known value (Figure 4-18).
Summary of Contents for LMX200 Series
Page 1: ...2015 00158 09 ...
Page 2: ...ii ...
Page 4: ...Product Registration iv ...
Page 10: ...x ...
Page 14: ...xiv ...
Page 48: ...Tools Setup 48 ...
Page 52: ...GPR Concepts 52 ...
Page 86: ...SliceView 86 ...
Page 96: ...Capturing Screens E mailing Mini Reports 96 Figure 10 3 Example of a Mini Report ...
Page 104: ...Transferring Data to a PC 104 ...
Page 108: ...Troubleshooting 108 ...
Page 111: ...Care Maintenance 111 ...
Page 112: ...Care Maintenance 112 ...
Page 114: ...Parts List Accessories 114 ...
Page 116: ...Parts List Accessories 116 ...
Page 118: ...Technical Specifications 118 ...
Page 120: ...Appendix A Health Safety Certification 120 ...
Page 128: ...Appendix B GPR Emissions Interference and Regulations 128 ...
Page 130: ...Appendix C Instrument Interference 130 ...
Page 132: ...Appendix D Safety around Explosive Devices 132 ...