Qty
Item
P/N
1
Orange fluorescent stick to test chlorophyll
output
MAA-520001
1
Scotch-Brite
™
scrub pad to clean copper
GXA-ZX0203
1
Ecominder
®
anti-fouling paint kit
BAA-003000
1
3/8" right-angle hose barb
RXA-031630
1
50 ml plastic bottle of Triton
®
X-100
50091
1
CT-DO cleaning device
50431
1
6-32 x 1/8" 316 stainless steel set screw for
Wiper (WQMx)
GXA-SX0074
3
8-32 x 5/16" flat head Phillips screws for
copper guard ring (WQMx)
GXA-SX0374
1
1/16" hex key for Wiper (WQMx)
GXA-SX0059
1
Wiper height gauge (WQMx)
MAA-230630
1
Blue fluorescent stick to test FDOM output
(WQMx)
MAA-520003
Installed parts
1
6-contact dummy connector
EXA-KX0038
1
4-contact dummy connector (digital external
data port)
EXA-KX0004
1
6-contact dummy connector (analog external
data port)
EXA-KX0183
2
Lock collars
EXA-KX0070
5.13 Bleach Injection System (BLIS) operation
The BLIS sends bleach into the pump manifold just below the DO sensor. The bleach
prevents biological fouling on the DO sensor membrane.
Adjust the BLIS for the type of environment it will be deployed in.
•
BLIS Operation
—Set to "On." When the conductivity value shows that the sensor is
in water, the BLIS sends bleach to the pump manifold after the first sample of every
hour.
•
Purge when in situ
—Set to "Yes." When the sensor is in situ, the BLIS pump
operates 20 times to remove air or water from the BLIS tubing. This makes sure a
sufficient quantity of bleach is available to protect the DO sensor.
•
BLIS Interval
—The BLIS sends the selected quantity of bleach after the first sample
of every hour.
•
BLIS Injection Volume
—Select the quantity of bleach to be used in 9 µL increments.
The manufacturer recommends a minimum of 36 µL.
•
BLIS Volume Used
—The host software makes a continuous estimate of the quantity
of bleach used in relation to the quantity selected by the user.
Reference