5.
Enter the serial number of the analog sensor in the SN: text box.
6.
Push
Edit
to change the pH Slope or pH Offset in the Analog Signal 1 area.
7.
Push
Done.
8.
Push
Edit
to change the ORP M or ORP B in the Analog Signal 2 area.
9.
Push
Done.
10.
Push
Send New Setup
.
The Status area will show "Loading..." until the new setup is stored in the sensor.
11.
Push
Done/Close Tab
.
12.
Push
Stop WQM Sample
, then
Yes, Stop WQM Sampling
.
13.
Remove power to the WQM.
The sensor is ready to be deployed.
3.4.3 Custom EDP set up
Use the "Custom Setup" function to set up sensors that can collect more than a single
"Signal" or parameter of data.
1.
Make sure that the external sensor is connected to the WQM.
2.
Make sure that the WQM sensor is in
In Standby
.
3.
Go to the
External Data Port
tab.
4.
Push
Get EDP Setup from WQM
.
The stored setup values will show in the window. If custom setup values do not show,
the user needs to enter them.
5.
Select "Custom Setup" in the Instrument Options drop-down menu.
6.
Do the procedures below to set up a three-parameter
ECO
sensor as in the example.
The user needs the sensor's device file to complete the setup.
Sensor set up and operation
14