3.
Lift the clamp up and off the CT and BLIS protective copper sleeves.
4.
Use a 5/32" hex key to remove the two 10-24 x ½" screws at the base of the BLIS
sleeve.
5.
Lift the copper BLIS sleeve up and off. Do not twist any more than necessary.
6.
If necessary, use a Scotch-Brite
™
scrub pad to remove any corrosion that "flakes" off.
7.
Flush the copper sleeve with clean water.
8.
Clean the intake area:
a.
Use a lint-free towel or tissue to clean the area.
b.
Do not scrape the inside the intake. Scratches may prevent the bleach reservoir
from a correct seal.
c.
Use the plastic-tipped syringe to flush the intake area with clean water.
4.5.1 Verify flow volume of BLIS
While the copper sleeve is still off of the intake area of the BLIS, use the 3 mL syringe
with the quick disconnect fitting to verify that the flow volume of the BLIS is correct.
1.
Connect the sensor to the PC and power supply.
2.
Start the software.
Sensor maintenance