3.8 Set up BLIS for deployment
D A N G E R
Bleach is corrosive and harmful to almost every part of the sensor. Make sure that bleach does
not get into the sensor.
The BLeach Injection System (BLIS) must be filled with bleach and its pump must be
primed before deployment. Only put bleach into a BLIS that will be deployed. If the sensor
will be put in storage, make sure there is NO bleach in the system.
Supplies:
User-supplied
Manufacturer-supplied
Clean water
5/32" hex key
Container for up to 150 ml of bleach
Plastic-tipped syringe (10 ml)
Household bleach
Syringe with quick-disconnect attachment (3 ml)
1.
Insert the long end of the 5/32" hex key straight through the hole in the copper BLIS
guard.
The key touches the reservoir release button.
2.
Push firmly on the key to release the reservoir.
3.
Insert the short end of the 5/32" hex key into one of the holes at the top of the
reservoir and lift the reservoir straight up and out.
Sensor set up and operation
18