User Manual SMO1000A-S
21/176
6.
Use the same method to test the other channels. If the square waveforms actually
shown do not match that in the figure above, please perform
“Probe Compensation”
in the next section.
WARNING
To avoid electric shock during the use of probe, please make sure that the
insulated wire of the probe is in good condition and do not touch the metallic
part of the probe when the probe is connected to high voltage source
Probe Compensation
When the probes are used for the first time, you should compensate the probes to match
the input channels of the oscilloscope. Non-compensated or poorly compensated probes
may cause measurement inaccuracy or error. The probe compensation procedures are as
follows.
1.
Set the switch to 10X on the probe.
2.
Perform steps 1, 2, 3 and 4 of
―
Function Inspection
ǁ
in the previous section.
3.
Check the waveforms displayed and compare them with the following:
Over
Perfectly
Under
Compensated
Compensated
Compensated
4.
Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole
on the probe until the waveform displayed is as the
―
Perfectly compensated
ǁ
in the
figure above.
Summary of Contents for SMO1000A-S
Page 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Page 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Page 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Page 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Page 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Page 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...