User Manual SMO1000A-S
122/176
To Make Automatic Measurement
Perform the steps below and select voltage or time parameters to make automatic
measurement.
1.
Press the
Measure
button on the front panel to enter the MEASURE function menu.
At the same, the frequency and period are enabled with the current trigger channel,
the statistics also was enabled.
2.
Press the
Source
softkey, and then use the
Universal Knob
to select the desired
channel. Only analog channels (CH1 and CH2) that are displayed are available for
measurements.
3.
To select and display measurement parameters. Press the
Type
softkey, and then
turn the
Universal Knob
to select the desired measurement parameter.
4.
Press the
Universal Knob
to add the measurement parameter, the parameters and
value will be shown above the menu, and the statistics status will update.
5.
To turn off the statistic function, press the
Statistics
softkey to select
Off
.
The measurement display area can display 5 measurement parameters at most, and the
measurements will arrange according to the selecting order. If add a sixth measurement
parameter, it will delete the first measurement.
Figure 64 Select the Measurement Parameter
Summary of Contents for SMO1000A-S
Page 1: ...Digital Oscilloscope SMO1000A S User Manual ...
Page 50: ...User Manual SMO1000A S 50 176 Figure 12 x Interpolation Figure 13 Sinx Interpolation ...
Page 76: ...User Manual SMO1000A S 76 176 Figure 28 Relative Window Trigger ...
Page 78: ...User Manual SMO1000A S 78 176 Figure 29 Interval Trigger ...
Page 82: ...User Manual SMO1000A S 82 176 Figure 32 Runt Trigger ...
Page 96: ...User Manual SMO1000A S 96 176 Figure 44 SPI Trigger ...