67 | 8 4
The instrument will automatically sample the standard capacitance value after impedance
transformation of the current.
FUNC:SOURce:STEP:OS:STAND
To set /inquiry about the STANDARD capacitance for OS
test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:OS:STAND<standard capacitance >
Query message: FUNC:SOUR:STEP <sn>:OS: STAND?
--
Data< standard capacitance >:
Data format: float
Data range: 0.001~40.000
Data accuracy: 0.001
Data unit: nF
--
Example:
Set the standard capacitance for OS test in STEP1 to 10nF.
Command message: FUNC:SOUR:STEP 1:OS:STAND 10
Query message: FUNC:SOUR:STEP 1:OS: STAND?
Return value: 10.000
5.2.3.8 Commands for BREAKDOWN VOLT Function
FUNC:BREAKdown:STEPMODE
To set /inquiry about the STEP MODE for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:STEPMODE<step mode >
Query message: FUNC: :BREAK: STEPMODE?
--
Data<step mode>:
Data format: integer
Data range: 0 ~ 1 (0: normal step; 1: voltage step)
--
Example:
Set the step mode for breakdown test to 1 (voltage step).
Command message: FUNC:BREAK:STEPMODE 1
Query message: FUNC:BREAK: STEPMODE?
Return value: 1
FUNC:BREAKdown:STEP
To set /inquiry about the STEP for voltage breakdown test.
--
Syntax:
Command message: FUNC:BREAK:STEP<step>
Query message: FUNC: :BREAK: STEP?
--
Data<step>:
Data format: integer
Data range: 2~999
Data accuracy: 1