29 | 8 4
MeasRang
: Select the appropriate current range for
IR
, calculate the current value according to the
test voltage and the insulation impedance of the DUT, ie I = U/R, and then select the appropriate
current range accordingly. Relationship between the current range and resistance measurement
range is shown in table
4-6.
Current Range
Resistance Measurement Range
Set Voltage
50V
~
499V
Set Voltage
500V
~
5000V
10mA(3
~
10mA)
0.1M
Ω
~
1M
Ω
0.1M
Ω
~
4.5M
Ω
3mA(0.3
~
3mA)
0.5M
Ω
~
4.5M
Ω
3.0M
Ω
~
15.0M
Ω
300uA(30
~
300uA)
3.0M
Ω
~
15.0M
Ω
10.0M
Ω
~
45M
Ω
30uA(3
~
30uA)
10.0M
Ω
~
45M
Ω
35.0M
Ω
~
450M
Ω
3uA(0.3
~
3uA)
45M
Ω
~
0.45G
Ω
0.40G
Ω
~
4.5G
Ω
300nA(20
~
300nA)
0.40G
Ω
~
4.9G
Ω
4.0G
Ω
~
50.0G
Ω
Table 4-6
4.3.4 Pause Mode
Figure 4-15 Pause Mode Setup
Message: message prompt string, you can enter English letters, Arabic numerals and
〖
.
〗
and
〖
-
〗
,
you can enter up to 16 characters.
PausTime: the action mode of setting the pause mode
a)
When set to continue test, the ending of the pause mode would be either the
【
START
】
key on
the front panel is pressed, or the external
【
START
】
signal is retriggered.
b) When set to 0.3~999.0s, if the set time finishes, the pause mode ends.