25 | 8 4
4.3.1 AC Withstanding Voltage Test Mode
Figure 4-12 Meas Mode Setup
Setting Item Range
Default
Explanation
MeasVolt
0.050
~
5.000kV,
0 — CLOSE
CLOSE
Set the voltage required for AC
withstanding voltage testing.
Ac Freq
50Hz or 60Hz
50Hz
Select
the
frequency
for
AC
withstanding voltage testing.
Hi Limt
0.001
~
120.0mA
500.0uA
Set the high limit of the test current. The
voltage is
≤
4kV, the current can reach
120mA; the voltage is >4kV, the current
is up to 100mA.
Low Limit
0.001mA
~
Hi
Limit,
0 — CLOSE
CLOSE
Set the low limit of the test current. The
low limit must be
≤
high limit or Close.
Arc Limit
1mA
~
20mA,
0 — CLOSE
CLOSE
Set the allowable AC arc current high
limit.
CLOSE means no requirement for arc.
Ramp Time
0.1
~
999.0s,
0 — CLOSE
CLOSE
Set the time required to rise to the set
voltage.
Close the ramp time, the voltage ramp is
completed during the first cycle of the
test.
MeasTime
0.3
~
999.0s
0 — Continue
3.0S
Set the measure time for the AC
withstanding voltage testing.
Set to 0 means that the test is continued
until the STOP key is pressed to end.
F1
F2
F3
F4
F5