82 | 8 4
Chapter 6 Technical Parameter
Model
SME1190
SME1190A
Withstanding Voltage Test
Output Voltage
AC
Range
0.05-5.0kV
Waveform
50/60Hz ±0.1% Sinusoidal Wave
Output
Power
500VA (5.0kV 100mA)
DC
Range
0.05-6.0kV
Output
Power
150VA (6.0kV 25mA)
Load Regulation
± (1%set value + 0.2%full scale) (rated power)
Voltage Resolution
2V
Voltage Accuracy
± (1%set value + 0.1%full scale)
Current Test Range
AC
Range
0.001mA-120mA (voltage
≤
4kV)0.001mA-100mA(voltage>4kV)
Resolution
0.001mA
Accuracy
± (1%r 0.5%full scale)
DC
Range
0.0001mA-25mA(voltage
≥
1.5kV)
0.0001mA-20mA(voltage
<
1.5kV)
Resolution
0.1uA
Accuracy
± (1%r 0.5%full scale)
Maximum Short-Circuit Current
200mA (AC test only)
Fast Discharge Function
Auto discharge after test ends (DCW)
Insulation Resistance Test
Output Voltage
DC:0.05-5.0kV
Voltage Resolution
2V
Voltage Accuracy
± (1% r 0.1% full scale)
Resistance Test Range
0.1M
Ω
– 50.0G
Ω
Resistance Test Accuracy
voltage
≥
1kV
1M
Ω
–1G
Ω
±(3%0.1%full scale)
1G
Ω
–10G
Ω
±(7%2%full scale)
10G
Ω
–50G
Ω
±(10%1%full scale)
voltage500V-1kV
0.1M
Ω
–1G
Ω
±(3%0.1%full scale)
1G
Ω
–10G
Ω
±(7%2%full scale)
10G
Ω
–50G
Ω
±(10%1%full scale)
voltage
<
500V
0.1M
Ω
–1G
Ω
±(5%2%full scale)
Fast Discharge Function
Auto discharge once the test ends
ARC Detection
Program Setting
AC
1.0mA-20.0mA
DC
1.0mA-10.0mA