![Scientific SME1190 User Manual Download Page 55](http://html1.mh-extra.com/html/scientific/sme1190/sme1190_user-manual_1217340055.webp)
55 | 8 4
5.2.1 SCPI Commands
Subsystem commands of SME1190/9110A:
●
DISPlay
●
FUNCtion
●
SYSTem
●
MMEM
●
FETC
5.2.2 DISPlay Subsystem Commands
DISPlay subsystem commands are mainly used to set the instrument display page.
The :DISPlay? query returns to the current page.
DISPlay
:
:
:
:
PAGE
Command Syntax: DISPlay:PAGE <page name>
<page name> are as follows:
TEST
Set the display page to: measurement display page
(TEST)
SETUP
Set the display page to: Measurement Setting
Page(SETUP)
SYST
Set the display page to: System Setting Page (SYST)
FILE
Set the display page to: File List Page (FILE)
MAIN
Set the display page to: Standby Home Page
Character? can query the current page.
--Example:
Set the display page to measurement display.
Command Syntax: DISP: PAGE TEST
Query syntax: DISPlay:PAGE?
Return format: TEST
Set measurement page display mode: 0, step mode; 1, list mode
Command Syntax: DISP:MODE 0
Query Syntax: DISP:MODE?
Return Format: 0
5.2.3 FUNCtion Subsystem Commands
5.2.3.1 FUNCtion Subsystem Commands
FUNCtion subsystem command is mainly used to set the test parameters of test function.
5.2.3.2 PROG Fuction Commands
FUNC:SOURce:STEP_<sn>:INS Add a new test project to the existing test plan (STEP).
FUNC:SOURce:STEP_<sn>:DEL In the existing test plan (STEP), delete the current test project.
FUNC:SOURce:STEP_<sn>:NEW Create an empty test programm to write a brand new test
program.