62 | 8 4
Data range: 0, 1.0
~
10.0mA (0 is OFF)
Data accuracy: 0.1mA
Data unit: mA
--
Example:
Set the ARC upper current for DCW test in STEP1 to1mA
Command message: FUNC:SOUR:STEP 1:DC:ARC 1
Query message: FUNC:SOUR:STEP 1:DC:ARC?
Return value: 1.0
FUNC:SOURce:STEP:DC:RAMPARC
To set /inquiry about the RAMP ARC for DCW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:DC:RAMPARC<current>
Query message:
FUNC:SOUR:STEP <sn>:DC:RAMPARC?
--
Data<voltage>:
Data format: float
Data range: 0, 1.0
~
10.0mA (0 is OFF)
Data accuracy: 0.1mA
Data unit: mA
--
Example:
Set the RAMP arc for DCW test in STEP1 to 1mA.
Command message: FUNC:SOUR:STEP 1:DC:RAMPARC 1
Query message: FUNC:SOUR:STEP 1:DC:RAMPARC?
Return message: 1.0
FUNC:SOURce:STEP:DC:RAMPARC
To set /inquiry about the RAMP state for DCW test.
--
Syntax:
Command message:
FUNC:SOUR:STEP <sn>:DC:RAMP<ON/OFF>or<1/0>
Query message:
FUNC:SOUR:STEP <sn>:DC:RAMP?
--
Data<voltage>:
Data format: character
Data range: OFF(0), ON(1)
--
Example:
Set the RAMP state for DCW test in STEP1 to ON.
Command message: FUNC:SOUR:STEP 1:DC:RAMP ON
Query message: FUNC:SOUR:STEP 1:DC:RAMP?
Return message: 1