75 | 8 4
Query message: SYSTem:MEA:HARDAGC?
Return value: ON
SYSTem:MEA:SOFTAGC
To set /inquiry about SOFTWARE COMPENSATION status.
--
Syntax:
Command message: SYSTem:MEA:SOFTAGC <ON/OFF>or<1/0>
Query message: SYSTem:MEA:SOFTAGC?
--
Data<ON/OFF>:
Data format: characters
Data range: OFF (0), ON (1)
--
Example:
Set the software compensation to ON.
Command message: SYSTem:MEA:SOFTAGC ON
Query message: SYSTem:MEA:SOFTAGC?
Return value: ON
SYSTem:MEA:AUTORANGE
To set /inquiry about AUTO RANGE status.
-
Syntax:
Command message: SYSTem:MEA:AUTORANGE <ON/OFF>or<1/0>
Query message: SYSTem:MEA:AUTORANGE?
--
Data<ON/OFF>:
Data format: characters
Data range: OFF (0), ON (1)
--
Example:
Set the auto rang to ON.
Command message: SYSTem:MEA:AUTORANGE 1
Query message: SYSTem:MEA:AUTORANGE?
Return value: 1
SYSTem:MEA:GFI
To set /inquiry about the state for GFI.
--
Syntax:
Command message: SYSTem:MEA:GFI <ON/OFF/FLOAT>or<1/0/2>
Query message: SYSTem:MEA:GFI?
--
Data<ON/OFF/FLOAT>:
Data format: characters
Data range: OFF (0), ON (1), FLOAT (2)
--
Example:
Set GFI to 1.
Command message: SYSTem:MEA:GFI 1
Query message: SYSTem:MEA:GFI?
Return value: 1