21 | 8 4
b)
When the electric shock protection is set to FLOAT, the high-voltage output high terminal HV1,
low terminal HV2 are in a float state, as shown in Figure 4-6. When the human body
accidentally touches the test high voltage terminal, the low terminal HV2 of the high voltage
circuit is not grounded ,but isolated from the ground, equivalent to connecting a high insulation
resistance, therefore, the current through the human body will not exceed 3 mA, thus ensuring
the safety of operator. As shown in Figure 4-7.
Figure 4-6
Figure 4-7 Relationship of HV1/HV2 Terminal to the Ground
4.2.2 System Environment Parameters Setting
Operation Instructions:
1.
Press
〖
SYSTEM
〗
to enter the system setting interface shown in Figure 4-2.
2.
Press [F2] key to enter the environment-related system settings shown in Figure 4-8.
3.
Press
〖
▲
〗
,
〖
▼
〗
to move the cursor to the parameter to be set. Change parameter settings
via F1~F6 keys or numeric keys.
The environment related parameter settings in the System settings are shown in Figure 4-8: