24 | 8 4
4.3 Meas SETUP
Operation Instructions:
1.
Press
〖
SETUP
〗
to enter the measurement setup interface shown in Figure 4-11.
2.
INSERT, DELETE, NEW, SAVE or select up or down by pressing the F1~F6.
3.
Press the right arrow key to move the cursor to the test mode option, as shown in Figure 4-12,
uses the F1~F5 keys to select whether the test mode is AC withstand voltage, DC withstand
voltage, insulation resistance, pause mode or open short circuit detection OSC.
4.
Move the cursor to the test parameter in the corresponding mode by using the arrow keys and
change the parameter setting by using the F1~F6 keys or numeric keys.
Figure 4-11 Meas Setup
Test procedures: current setting step number / total number of test steps. The change of the test
steps are as shown in the table below:
Key Function
Explanation
F1
INSERT
Add a new test item. The current item and subsequent items
will move one step backward.
F2
DELETE
Delete the current test item. The subsequent item will move
one step forward.
F3
NEW
Create a blank test program (PROG). Please remember to
save the test program after finishing writing.
F4
SAVE
Save the current test file
F5
Up
Move to the display screen of the previous step of the current
step.
F6
Down
Move to the display screen of the next step of the current step.
Table 4-3
F1
F2
F3
F4
F5
F6