81
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Technical Data (cont.)
Myogram Filter (muscle tremor filter):
25 Hz or 35 Hz, programmable (not active on
averaged waveform). The stored ECGs can be printed
with or without filter.
Line Frequency Filter:
Distortion-free suppression of superimposed 50 or 60 Hz
sinusoidal interferences.
Patient Input:
Fully floating and isolated, defibrillation protected.
Patient Leakage Current:
< 5 µA
Safety Standard:
CF according to IEC 60601-1 and IEC 60601-2-25
Safety Class:
I according to IEC 60601-1 (with internal power supply)
Environmental Conditions:
Temperature, Operating:
10
O
to 40
O
C
Temperature, Storage:
-10
O
to 50
O
C
Relative humidity:
25 to 95%
(non condensing)
Atmospheric pressure:
700 to 1060 hPa
Control Panel:
Rubber keys
AT-1 smartprint only :
External Printer :
12 channel high resolution print-out, standard
Centronics connecting cable
Available Configurations
The AT-1 / AT-1smartprint is available in three versions:
Standard Version:
Unit with ECG recording and printout capabilities.
Version M:
Unit with additional ECG measurement program.
Version C:
Unit with additional ECG Interpretation program
(including measurements).