Control and Power Supply Board MK 11-10
The Control and Power Supply Board MK 11-10 is secured to the base assembly and
located under the ECG Interface Board MK 11-2.
Prerequisites
• The Warnings and Cautions at the beginning of the chapter must be observed.
• The top housing must be removed and all external cable assemblies
disconnected.
• The ECG Interface Board must be removed.
• The printer / paper tray assembly must be removed.
Tools and Equipment
Posi-drive screwdriver
Parts
Control and Power Supply board MK 11-10. Part number as detailed in Chapter 6.
Board Removal
WARNING
ENSURE THAT THE MAINS CABLE IS DISCONNECTED BEFORE COMMENCING
To remove the Control and Power Supply Board MK 11-10 proceed as follows:
1. Disconnect the following connectors:
•
live and neutral bayonet connectors to the mains connector
•
battery connectors
2. Unscrew the 12 fixing screws (four on the mains transformer) and lift out the board.
red (live)
black (neutral)