SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Automatic Mode (ECG) Settings (cont.)
Selecting Rhythm Leads
The rhythm leads are printed out as defined. Two separate rhythm leads can be selected.
The following formats can be set:
Rhythm Leads (interpretation option only)
Entry Key
Sequence
Setup Format
ALT
3
Define rhythm lead one
4
Define rhythm lead two
The 2 rhythm leads are defined as follows:
Extremity Leads
Entry Key Sequence
Lead
Confirm
ALT
3 or 4
8
1
I
Press
STOP key
2
II
3
III
4
aVR
5
aVL
6
aVF
Precordial Leads
Entry Key Sequence
Lead
Confirm
ALT
3 or 4
9
1
V1
Press
STOP key
2
V2
3
V3
4
V4
5
V5
6
V6
Confirm the selection by pressing the STOP key