29
SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Automatic Mode (ECG) Settings, Internal Printer (cont.)
Measurements and Markings (M and C versions only)
To define the measurements and markings proceed as follows:
Measurements (M and C Options Only) Internal printer
Entry Key Sequence
Printout
Confirm
ALT
1 or 2
3
5
Detailed table of measurement results
omitted - however, the values of electrical
axes, intervals, and heart rate are not
suppressed.
Press STOP
key
6
Detailed table of measurement results is
printed
7
Referenece markings are omitted
8
Reference markings (beginning and end of
P wave and QRS, and end of T wave)
are added to the ECG average cycles
Interpretation (C version only)
To print or suppress interpretation statements on the printout proceed as follows:
Interpretation (Interpretation Option Only) Internal printer
Entry Key Sequence
Printout
Confirm
ALT
1 or 2
4
5
Interpretation is omitted
Press STOP
key
6
Interpretation is printed
Confirm the selection by pressing the STOP key
Full details of the interpretation option are given in the SCHILLER ECG Measurement and
Interpretation booklet (art. No. 2.510 179).