SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Settings (cont.)
The defined formats and settings for your unit can be checked as follows:
ALT - 0 - 1 - any number
A printout of the defined settings will be produced and gives the following information,
depending on the installed software.
Unit designation
Software version, Software option installed (C =
Interpretation) and interpretation version
Serial number
Serial number of the unit
Leads
Standard (S) or Cabrera (C)
ECG Format
Long (ooo), Short (o) or Suppressed (-)
M ECG
Average cycles as defined in auto ECG recording setup
(e.g. 4 * 3 (25 mm/s) + 2)
Measurements
Enabled (+) or Suppressed (-)
Marks
Enabled (+) or Suppressed (-)
Interpretation
Enabled (+) or Suppressed (-)
Selected Rhythm leads
Leads selected for R1 and R2
Automatic Centering
Enabled (+) or Suppressed (-)
Paper type
Z-folded
Printout of signals
Sequential or Simultaneous
Baseline Filter
0.05, 0.15 or 0.30 Hz
Mains Filter
50, 60 Hz or OFF (-)
Myogram Filter
25 or 35 Hz, ON (+) or OFF (-)
Interpretation settings:
N/A:+/-
‘normal/abnormal’ is written (+) or
suppressed (-)
U:+/-
‘unconfirmed report’ is written (+) or
suppressed (-)
A30:+/-
patient age is assumed to be < 30 (-)
or >30 (+)
S: +/-
low (-) or high (+) sensitivity