SCHILLER AT-1 / AT-1smartprint ECG Unit
SERVICE HANDBOOK Issue b June 2001
Art.No. 2.540022
Settings (cont.)
Myogram Filter
The Myogram filter suppresses disturbances caused by strong muscle tremor. The set
value will be the new upper limit of the frequency range as soon as the FILTER key is
pressed on or programmed as default when the unit is switched on. When the Myogram
filter is on, the value, i.e. 35 Hz is displayed on the bottom line of the printout.
Myogram Filter
Entry Key Sequence
Setting
Confirm
ALT
8
2
Myogram Filter 25 Hz
Press STOP
key
3
Myogram Filter 35 Hz
1
Myogram Filter active when the unit is first
switched on (marked on printout with +)
8
Myogram Filter off when the unit is first
switched on (marked on printout with -)
Confirm the selection by pressing the STOP key
The myogram filter is switched on and off manually with the FILTER KEY
Note:
An ECG recorded in auto mode is stored unfiltered. It is therefore possible to
print the stored ECG either with or without passing the myogram filter. Filter
ON is indicated in the bottom information line of the printout. When the FILTER
key is pressed again, the filter is switched off and the `35 Hz` indication on
the bottom information line of the printout is removed. The cutoff frequency
of the myogram filter is set to either 25 or 35 Hz.