![Schiller CARDIOVIT AT-1 Service Handbook Download Page 80](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075080.webp)
80
Technical Data
Technical data subject to change without notice.
Dimensions:
290 x 210 x 69 mm
Weight:
2.9 kg
Mains Supply:
100 to 115 / 220 to 240 VAC, 50/60 Hz
Battery:
Built-in 12 V lead-acid battery (rechargeable)
Battery Capacity:
2 hours normal use
Power Consumption:
Recording: 28 VA max
Leads:
Standard / Cabrera
Paper Speed:
5 / 25 / 50 mm/s (direct)
Sensitivity:
5 /10 / 20 mm/mV, either automatically adjusted or
manually selected
Chart Paper:
Thermoreactive - Z-folded, 90 mm wide, perforation 90
mm
Printing Process:
High-resolution thermal printhead, 8 dots per mm
Recording Tracks:
3 channels, positioned at optimal width on 80 mm,
automatic baseline adjustment
Automatic Lead Programs:
3 channel representation of 12 simultaneously acquired
standard leads
Data Record:
Listing of ECG recording data
Version C: ECG measurement results (intervals,
amplitudes, electrical axes), Sokolow Index, average
complexes with optional measurement reference
markings, and interpretation.
ECG Storage:
Circular input memory for 10 s, 12-lead ECG.
Frequency Range
of Digital Recorder:
0 to 150 Hz (IEC)
0 to 150 Hz (AHA)
ECG Amplifier:
Simultaneous, synchronous registration of all 9 active
electrode signals (= 12 standard leads)
Sampling frequency:
1000 Hz
Digital resolution:
5 µV
Dynamic range:
± 10 mVAC
Max. electrode potential:
±300 mVDC
Time constant:
>3.2 s
Frequency response:
0.05 to 150 Hz (-3 dB)
Input impedance:
> 10 MOhms