![Schiller CARDIOVIT AT-1 Service Handbook Download Page 50](http://html1.mh-extra.com/html/schiller/cardiovit-at-1/cardiovit-at-1_service-handbook_1212075050.webp)
50
Printer Check
External Printer Problems
Possible Printer Problem
External Inkjet printer
Corrective Action
No printout
Check that the USB and the Centronics interface
on the external printer are not simultaneously
engaged.
Poor quality printout.
Make sure that the printer and the AT-1
smartprint are properly grounded via the
potential equalisation stud.
No printout
No format has been selected in Settings. Select
a proper format.