Automatic Mode (ECG) Settings, Internal Printer
Two separate Auto formats can be defined for the internal printer in the AT-1 /
AT-1smartprint. When defining auto format 1 the key sequence
ALT - 1
precedes the setting. When defining auto format 2 the key sequence
ALT - 2
precedes the setting.
The automatic mode formats are detailed on the following pages. The ECG format is set
as follows:
ECG Format Internal Printer
Entry Key Sequence
Printout
Confirm
ALT
1 or 2
1
5
No leads printed
Press STOP key
6
Leads are printed in short form (3 sheets)
7
Leads are printed in long form (6 sheets)
8
Chart Speed 25mm/s
9
Chart Speed 50mm/s
Average Cycles
The Average cycles are defined as follows:
Note: Lead selection for the rhythm lead(s) is defined on page 33.
Average Cycles (interpretation option only) Internal printer
Entry Key Sequence
Printout
Confirm
ALT
1 or 2
2
5
No average lead cycles are printed
Press STOP
key
6
4 x 3 (25 mm/s). The average
complexes are printed in 4 groups of
three leads at a chart speed of 25mm/s
7
4 x 3 (50 mm/s). The average
complexes are printed in 4 groups of
three leads at a chart speed of 50mm/s
8
6 x 2 (50 mm/s) + 1 rhythm lead
(25mm/s). The average complexes are
printed in 6 groups of two leads at a
chart speed of 50mm/s
9
12 x 1 (25 mm/s) + 2 rhythm leads
(25mm/s). The average complexes are
printed out for all 12 leads at a chart
speed of 50mm/s