Tel: 886. 909 602 109 Email: [email protected]
www.salukitec.com
84
Fig.66 The Configuration Area of Output Channel
Table 10. Output Channel Configuration Options
Channel 1
Choose CH1 to output the required values. The value types contain
the X/Y/R/θ value of signal and signal harmonic, the noise value and
the E1/E2/E3/E4 value.
Channel 2
Choose CH2 to output the required values. The value types contain
the X/Y/R/θ value of signal and signal harmonic, the noise value and
the E1/E2/E3/E4 value.
Offset (%)
The range is -100% to +100%, the minimal step is 0.01%, 0.00% in
default. It can only set the R/X/Y value.
Expand
The range is 1-256, 1 in default.
Speed
Choose ‘Fast’/’Slow’. When choosing ‘Fast’, the CH1 and CH2 can
only output the R/X/Y value.
Note that:
After setting the CH1 or CH2 output configuration, click the button
to finish the configuration.
All settings can be changed simultaneously and you just need to click this button once.
After changing the settings, you need to click
to finish this configuration. Otherwise, changes
of settings will not be saved.
5.2.5 Equation and System Configuration