the [Auto Reserve] key. Auto Reserve may take some time. The parameter i represents
the selected channel (A or B).
APHSD i
The APHSD command performs the Auto Phase function. It is the same as pressing the
[Auto Phase] key. The parameter i represents the selected channel (A or B).The outputs
will take some time to reach their new values. Do not send a new APHSD command
without waiting the appropriate amount time. If the phase is unstable, then APHSD will
do nothing.
Use PHAS? Command to query the new value of the phase shift and see if APHSD
changed the phase shift.
ASCLD i
The ASCLD command performs the Auto Scale function. It is the same as pressing the
[Auto Scale] key. The parameter i represents the selected channel (A or B).
4.2.8 Data Store Commands
SRATD(?)i{,x}
The SRATD command sets or queries the step time of data sampling. The parameter i
represents the selected channel (A or B). SRATD? 1 command will queries channel A for
the value of step time in sampling. The setting range of i is 1ms-100,000ms. The
minimum resolution is 1ms.
SLEND(?)i{,j}
The SLEND command sets or queries the length of data sampling. The parameter i
represents the selected channel (A or B). The parameter j indicates the length of data
sampling. The maximum length is 16384.
SSLED(?)i,j{,k}
The SSLED command sets or queries the chosen parameters in four buffers. The
parameter i represents the selected channel (A or B). The parameter j indicates the chosen
parameters. The parameter i must be set when sending this command. The parameter i
selects <Buffer 1> (k=1), <Buffer 2> (k=2), <Buffer 3> (k=3) or <Buffer 4> (k=4). The
parameter j means the recorded measurement value in buffers.
The parameter k selects the parameters as the following.
k
Parameter
0
R
1
X
2
Y
3
θ
4
Rh1
5
Xh1
6
Yh1
7
θh1
8
Rh2
9
Xh2