(1) Press [INPUT/FILTERS] and then set the input mode to <I> and <Amplitude: 1M>. Other settings remain the
default.
(2) Press [REF/PHASE] and then set <Ref. Source: External> and <Type: TTL>.
(3) Press [GAIN/TC] and then set <Sensitivity: 1μA> which can be adjusted in measurement, <Reserve: Normal>,
<TC: 300ms> and <Roll off: 12dB/oct>. Finally, turn off the SYNC filter.
(4) Press [OUTPUT/OFFSET] and then set <CH1>, <Source: R> and <Speed: Fast>.
3. Build up the spectral measurement platform.
Use a BNC cable to connect the output of the photoelectric detector and the IN+ input of the SE1022D.
Use a BNC cable to connect the OUTPUT of the optical chopper and the REF IN of the SE1022D.
Use a BNC cable to connect the data collection platform and the CH1 OUT of the SE1022D. The sketch is shown
in Fig.96.
Reference Signal
Chopper
Grating
Monochrometer
Photoelectricity
Sensor
SE1022D
Light Source
Control System of
Monochrometer
Data
Acquisition
Platform
Serial Communication
PC
Fig.96 Spectral Measurement Platform Diagram
Fig.97 Spectral Measurement Platform Physical Connection