4: 300 us
13: 10 s
5: 1 ms
14: 30 s
6: 3 ms
15: 100 s
7: 10 ms
16: 300 s
8: 30 ms
17: 1000 s
9: 100 ms
18: 3000 s
OFSLD(?)i{,j}
The OFSLD command sets or queries the low pass filter slope. The parameter i represents
the selected channel (A or B). The parameter j indicates the low pass filter slope. The
parameter i selects <6dB/oct> (j=0), <12dB/oct> (j=1), <18dB/oct> (j=2) or <24dB/oct>
(j=3).
SYNCD(?)i{,j}
The SYNCD command sets or queries the status of synchronous filter. The parameter i
represents the selected channel (A or B). The parameter j indicates the status of
synchronous filter. The parameter i selects Off (j=0) or On (j=1). It is recommended to
turn on the synchronous filtering when the reference frequency is less than 200 Hz.
4.2.5 Displayed and Output Commands
FPOPD(?)j{,k}
The FPOPD i command sets and queries the output channel on the front panel of
SE1022D. The parameter j must be set when sending this command. The parameter j
selects CH1 (j=1) or CH2 (j=2) as the output channel. The parameter k is used to select
the type of output source. More details are as follow.
k
Parameter
0
A-R
1
A-X
2
A-Y
3
A-θ
4
A-Rh1
5
A-Xh1
6
A-Yh1
7
A-θh1
8
A-Rh2
9
A-Xh2
10
A-Yh2
11
A-θh2
12
A-Noise
13
A-E1
14
A-E2
15
A-E3