When selecting it, markers of all Traces will be displayed, and when clearing it, only markers of Active Traces will be displayed.
Marker type
[Normal]
Standard marker has fixed X-axis position, Y-axis position will change with amplitude of Trace data, and marker can be shifted left
and right on X-axis by changing stimulus value of marker.
[Fixed]
According to the position on Traces when it is set up as Fixed type marker, the marker will constantly keep fixed X-axis and Y-axis
coordinates unchanged; it will not fluctuate with change of Trace data amplitude. Its position on X-axis can be shifted by changing its
stimulus value, but coordinate on Y-axis can remain unchanged. This type of marker is mainly used to observe change of Trace
marker, for example, the fixed type marker can be used to compare variance of insertion loss before and after tuning wave-filter.
6.1.6.
Marker Table
Marker Table can be enabled to display all marker data of Active Traces, and marker data will be displayed in specified format of
each marker.
6.1.6.1.
Settings of show table
By mouse or touch screen -marker menu
1) Click [Marker/Analysis], point to [Marker] in Marker/Analysis menu.
2) Select [Show Table] in submenu to enable Show table.
3) Select [Hide Table] in submenu to close Show table.
Figure 6-9 Settings of Show table via Marker Menu
By mouse or touch screen -view menu
1) Click [Response], point to [Table] in [Display] menu.
2) Select [Marker Table] in submenu of Table to enable Marker Table display.
3) Click clear and select [Marker Table] in submenu of Table to close Marker Table display.