By front panel keys
1) Press
【
Meas
】
in RESPONSE area.
2) Press the corresponding soft-key to select S-parameter Measure type on existing soft-key menu.
4.2.4.3.
Changing current Active Trace to be arbitrary ratio type
By mouse or touch screen
1) Single click [Response] and point to [Measure] in Response menu to display submenu of Measure.
2) Single click menu item of [Measure…] in submenu to display Measure dialog box.
3) Single click [More Type] button to display Measure dialog box.
4) Select the source output port of Analyzer from [Source] box.
5) Click [Scale Type] to mark check box.
6) Select input receiver for ratio Measure from Input zone.
7) Select reference receiver for ratio Measure from Reference zone.
8) Single click [OK] button to close the dialog box after the setting is completed.
Fig. 4-7 Changing Current Active Trace to be Arbitrary Ratio Type
By front panel keys
1) Press
【
Meas
】
in RESPONSE area and single click the soft-key corresponding to [Measure…] to display Measure dialog
box.
2) Press
【
Tab
】
in ADJUST area to switch to [More Type] button and press
【
Click
】
to display Next-level Measure dialog
box.
3) Press
【
Tab
】
in ADJUST area to switch to [Source] box and press
【
↑
】
and
【
↓
】
or rotate knob to select source output
port of Analyzer.
4) Press
【
Tab
】
in ADJUST area to switch to [Scale Type] and press
【
Click
】
to mark check box.
5) Press
【
Tab
】
in ADJUST area to switch to Input zone and press
【
↑
】
and
【
↓
】
or rotate knob to select input receiver
of ratio Measure.
6) Press
【
Tab
】
in ADJUST area to switch to Reference zone and press
【
↑
】
and
【
↓
】
or rotate knob to select reference
receiver of ratio Measure.
7) Press
【
OK
】
in ENTRY area to close dialog box.